Class II B2 Biological Safety Cabinet
Class II B2 Biological Safety Cabinet
Class II B2 Biosafety Cabinet designed for total exhaust applications, featuring ULPA filtration, precise airflow control, and advanced safety systems for high-risk laboratory environments.
Description
| Specification | BSC-1100IIB2-X | BSC-1300IIB2-X | BSC-1500IIB2-X | BSC-1800IIB2-X |
|---|---|---|---|---|
| Internal Size (W×D×H) | 940 × 600 × 660 mm | 1150 × 600 × 660 mm | 1350 × 600 × 660 mm | 1700 × 600 × 660 mm |
| External Size (W×D×H) | 1100 × 750 × 2250 mm | 1300 × 750 × 2250 mm | 1500 × 760 × 2250 mm | 1873 × 775 × 2270 mm |
| Tested Opening | Safety height 200 mm (8″) | Safety height 200 mm (8″) | Safety height 200 mm (8″) | Safety height 200 mm (8″) |
| Max Opening | 420 mm (17″) | 420 mm (17″) | 500 mm (20″) | 480 mm (20″) |
| Inflow Velocity | 0.53 ± 0.025 m/s | 0.53 ± 0.025 m/s | 0.53 ± 0.025 m/s | 0.53 ± 0.025 m/s |
| Down flow Velocity | 0.33 ± 0.025 m/s | 0.33 ± 0.025 m/s | 0.33 ± 0.025 m/s | 0.33 ± 0.025 m/s |
| Pre-Filter | Washable | Washable | Washable | Washable |
| ULPA Filter | Two, 99.9995% efficiency at 0.12 μm, filter life indicator | Two, 99.9995% efficiency at 0.12 μm, filter life indicator | Two, 99.9995% efficiency at 0.12 μm, filter life indicator | Two, 99.9995% efficiency at 0.12 μm, filter life indicator |
| Front Window | Motorized, two-layer laminated toughened glass ≥5mm, anti UV | Motorized, two-layer laminated toughened glass ≥5mm, anti UV | Motorized, two-layer laminated toughened glass ≥5mm, anti UV | Motorized, two-layer laminated toughened glass ≥5mm, anti UV |
| Noise | NSF49 ≤61 dB / EN12469 ≤58 dB | NSF49 ≤61 dB / EN12469 ≤58 dB | NSF49 ≤61 dB / EN12469 ≤58 dB | NSF49 ≤61 dB / EN12469 ≤58 dB |
| UV Lamp | 30W × 1 | 30W × 1 | 40W × 1 | 40W × 1 |
| UV timer, UV life indicator, emission of 253.7 nanometers for most efficient decontamination. | UV timer, UV life indicator, emission of 253.7 nanometers for most efficient decontamination. | UV timer, UV life indicator, emission of 253.7 nanometers for most efficient decontamination. | UV timer, UV life indicator, emission of 253.7 nanometers for most efficient decontamination. | |
| Illuminating Lamp | LED Lamp | LED Lamp | LED Lamp | LED Lamp |
| 12W × 2 | 14W × 2 | 16W × 2 | 16W × 2 | |
| Illumination | ≥1000 Lux | ≥1000 Lux | ≥1000 Lux | ≥1000 Lux |
| Consumption | 700 W | 850 W | 900 W | 1200 W |
| Waterproof Sockets | Two, total load of two sockets: 500W | Two, total load of two sockets: 500W | Two, total load of two sockets: 500W | Two, total load of two sockets: 500W |
- Time reserve function.
- Motorized front window.
- ULPA filter life and UV life indicator.
- Automatic air speed adjustable with filter block.
- With memory function in case of power-failure.
- Work area surrounded by negative pressure, it can ensure maximum safety in work area.
- Most accessories are standard, which saves your money. There is no need to pay more.
- Audio and visual alarm (filter replacement, window over height, abnormal air flow velocity, etc).
- Remote control. All functions can be realized with it, making the operation much easier and more convenient.
- Interlock function: UV lamp and front window; UV lamp and blower, fluorescent lamp; blower and front window.
- Foot switch. Adjust front window height by foot during experiment, to avoid airflow turbulence caused by arm movement.
Request a Quote Today
Tell us your requirements and our team will get back to you with a tailored solution, pricing, and expert support.
Answers to Common Fume Hood Questions
Quick guidance to help you choose the right fume hood for your lab or facility.
What makes Nupon fume hoods different from others in the market?
Nupon fume hoods are built with cold-rolled steel exteriors, acid and alkali-resistant interiors, UV sterilization, and precision airflow controls. They combine durability, safety, and modern design at a competitive price — giving you big-brand quality with personalized support.
Do your fume hoods meet international safety standards?
Yes. All Nupon fume hoods are designed in compliance with international laboratory safety standards and undergo strict quality control to ensure maximum protection for operators, equipment, and the environment.
Are Nupon fume hoods customizable for specific laboratory needs?
Absolutely. We offer various sizes, airflow controls, and material options (such as chemical-resistant interiors and ergonomic slope-front designs) to suit different applications — from schools and universities to industrial facilities.
How do I maintain and clean a Nupon fume hood?
Our fume hoods are designed for easy maintenance. Regular cleaning with mild, approved cleaning agents and routine checks of the filters, airflow, and UV lamps will keep your unit performing at its best. We also provide guidance and after-sales support for proper care.
Does Nupon provide after-sales service and technical support?
Yes. We stand behind our products with dedicated after-sales service, spare parts availability, and technical support to ensure your equipment continues to deliver reliable protection long after installation.
What makes Nupon fume hoods different from others in the market?
Nupon fume hoods are built with cold-rolled steel exteriors, acid and alkali-resistant interiors, UV sterilization, and precision airflow controls. They combine durability, safety, and modern design at a competitive price — giving you big-brand quality with personalized support.
Do your fume hoods meet international safety standards?
Yes. All Nupon fume hoods are designed in compliance with international laboratory safety standards and undergo strict quality control to ensure maximum protection for operators, equipment, and the environment.
Are Nupon fume hoods customizable for specific laboratory needs?
Absolutely. We offer various sizes, airflow controls, and material options (such as chemical-resistant interiors and ergonomic slope-front designs) to suit different applications — from schools and universities to industrial facilities.
How do I maintain and clean a Nupon fume hood?
Our fume hoods are designed for easy maintenance. Regular cleaning with mild, approved cleaning agents and routine checks of the filters, airflow, and UV lamps will keep your unit performing at its best. We also provide guidance and after-sales support for proper care.
Does Nupon provide after-sales service and technical support?
Yes. We stand behind our products with dedicated after-sales service, spare parts availability, and technical support to ensure your equipment continues to deliver reliable protection long after installation.